Remote Commands
R&S
®
Spectrum Rider FPH
453
User Manual 1321.1011.02 ─ 11
Parameters:
<SearchLimit>
Sets the right marker search limit.
The unit depends on the measurement, e.g. Hz for measure-
ments in the frequency domain and seconds for measurements
in the time domain.
Range:
Maximum span.
*RST:
– (is set to the right diagram border when switching
on search limits)
Default unit: GHz
Example:
CALC:MARK:X:SLIM ON
CALC:MARK:X:SLIM:LEFT 10MHz
CALC:MARK:X:SLIM:RIGH 100MHz
Turns on search limits and defines a search range from 10 MHz
to 100 MHz.
CALCulate<1...2>:MARKer<1...6>:Y?
This command queries the absolute vertical position of a marker.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
The unit of the return value depends on
UNIT<1...2>:POWer
.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<MarkerPosition>
Numeric value of the marker position.
Example:
INIT:CONF OFF
CALC:MARK2 ON
Turns on single sweep mode and marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Queries the position of marker 2.
Usage:
Query only
Marker Functions
The following commands perform various kinds of analysis at the marker position.
List of Commands
●
CALCulate<1...2>:MARKer<1...6>:FREQuency:MODE
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation[:STATe]
Remote Control - Commands