Remote Commands
R&S
®
Spectrum Rider FPH
412
User Manual 1321.1011.02 ─ 11
Example:
CALC:DELT:MOD REL
Delta marker positions are relative to marker 1.
CALC:DELT2:X 10.7MHz
Positions delta marker 2 10.7 MHz to the right of marker 1.
CALC:DELT2:X?
CALC:DELT2:X:REL?
Queries the absolute and relative position of delta marker 2.
CALCulate<1...2>:DELTamarker<1...6>:X:RELative
<FrequencyOrTime>
This command positions a delta marker on a position relative to the reference marker.
If necessary, the corresponding delta marker is activated first.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Parameters:
<FrequencyOrTime> Defines the distance of the marker to the reference marker.
Range:
Depends on the current scaling of the horizontal
axis.
Default unit: GHz
Example:
CALC:DELT3:X:REL 5 kHz
Sets the delta marker at a distance of 5 kHz to the reference
position.
CALCulate<1...2>:DELTamarker<1...6>:Y?
This command queries the vertical position of a delta marker. The result is always a
relative value in relation marker 1.
If necessary, the corresponding delta marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
In spectrum analyzer mode, the unit depends on the unit you have set and the scaling
of the vertical axis.
Parameter or measuring functions
Output unit
DBM | DBPW | DBUV | DBMV | DBUA
dB (lin/log)
WATT | VOLT |AMPere
dB (lin), % (log)
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Parameters:
<MarkerPosition>
Remote Control - Commands