Remote Commands
R&S
®
Spectrum Rider FPH
410
User Manual 1321.1011.02 ─ 11
CALCulate<1...2>:MARKer<1...6>:X
<FrequencyOrTime>
This command positions a marker on a particular coordinate on the horizontal axis.
If one or more delta markers (2 to 6) are already active, the command turns these delta
markers into normal markers.
Note that it is possible to place the marker outside the visible trace. In that case, this
value is invalid.
If necessary, the corresponding delta marker is activated first.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<FrequencyOrTime> Indicates the coordinate on the horizontal axis.
The unit in spectrum analyzer mode depends on the measure-
ment, e.g. Hz for measurements in the frequency domain and
seconds for measurements in the time domain.
Range:
Maximum span.
Default unit: GHz
Example:
CALC:MARK2:X 10.7MHz
Positions marker 2 to frequency 10.7 MHz.
CALCulate<1...2>:MARKer<1...6>:Y?
This command queries the absolute vertical position of a marker.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
The unit of the return value depends on
UNIT<1...2>:POWer
.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<MarkerPosition>
Numeric value of the marker position.
Example:
INIT:CONF OFF
CALC:MARK2 ON
Turns on single sweep mode and marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Queries the position of marker 2.
Usage:
Query only
Remote Control - Commands
www.allice.de