Remote Commands
R&S
®
Spectrum Rider FPH
405
User Manual 1321.1011.02 ─ 11
With a reference level offset ≠ 0, the value range of the reference level is modified by
the offset. You can set the offset with
TRACe<1...2>:Y[:SCALe]:RLEVel:OFFSet
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLevel>
Sets the reference level; the unit depends on
Range:
Specified in the data sheet.
*RST:
–20 dBm
Default unit: dBm
Example:
DISP:TRAC:Y:RLEV –60dBm
Sets the reference level to -60 dBm.
INPut:ATTenuation<1...2>
<Attenuation>
This command defines the input attenuation.
The attenuation is coupled to the reference level. If you set the attenuation independ-
ently, the R&S Spectrum Rider turns off this coupling.
The R&S Spectrum Rider adjusts the reference level if it cannot be set for the current
RF attenuation.
Suffix:
<1...2>
.
1...2
RF port selection. Default selection at 1.
RF port 2 is only applicable for R&S ZNH.
Parameters:
<Attenuation>
Range:
0 dB to 40 dB
*RST:
0 dB (AUTO is ON)
Default unit: dB
Example:
INP:ATT1 30dB
Defines RF attenuation of 30 dB at RF port 1 and deactivates
coupling to the reference level.
INPut:ATTenuation:AUTO
<State>
This command couples and decouples input attenuation to the reference level.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
INP:ATT:AUTO ON
Couples the attenuation set on the attenuator to the reference
level.