Remote Commands
R&S
®
Spectrum Rider FPH
355
User Manual 1321.1011.02 ─ 11
<X0...X99>
Data points on the x-axis.
Note: a limit line may consist of up to 100 horizontal data points.
<Y0...Y99>
Data points on the y-axis.
Example:
CALC:LIM:DEF
'Line','Example',HZ,ABS,DBM,10000000,-10,
10000000,0,20000000,0
Defines a limit line with three data points.
Usage:
Setting only
CALCulate<1...2>:LIMit<1...2>:DELete
This command deletes a limit line.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Example:
CALC:LIM2:DEL
Deletes the second limit line.
CALCulate<1...2>:DLINe
<LinePosition>
This command defines the position of a display line.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<LinePosition>
Numeric value with a variable range and unit.
You can use any unit you want, the R&S Spectrum Rider then
converts the unit to the currently selected unit. If you omit a unit,
the R&S Spectrum Rider uses the currently selected unit.
*RST:
– (STATe to OFF)
Default unit: dBm
Example:
CALC:DLIN –20dBm
Sets the display line threshold to -20 dBm.
CALCulate<1...2>:DLINe:STATe
<State>
This command turns display lines on and off.