Remote Commands
R&S
®
Spectrum Rider FPH
349
User Manual 1321.1011.02 ─ 11
Example:
CALC:MARK3:FUNC:DEM ON
Switches on the demodulation for marker 3.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation:HOLD
<HoldTime>
This command defines the hold time at the marker position for the demodulation with
span > 0.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<HoldTime>
Range:
100 ms to 500 s
*RST:
0.5 (DEModulation is set to OFF)
Default unit: s
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Sets a hold time of 3 seconds.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation:SELect
<DemodType>
This command selects the type of demodulation type for the audio demodulator.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<DemodType>
AM | FM
*RST:
AM
Example:
CALC:MARK:FUNC:DEM:SEL FM
Selects FM demodulation.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown
<MarkerDistance>
This command defines the distance of the n dB down markers to the reference marker.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<MarkerDistance>
Distance of the temporary markers to the reference marker in
dB.
*RST:
3 dB
Default unit: dB
Example:
See
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown:FREQuency?
This command queries the horizontal position of the n dB down markers.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH