Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
193
User Manual 1321.1011.02 ─ 11
6.2.6.3
Working with a Second Trace
In spectrum mode, you can use up to two traces. All two traces are based on the same
settings, except the trace settings like the trace mode or the detector. You can use the
second trace to compare, for example, four different detector settings.
In the default state, only trace 1 is active.
1. Press [TRACE] key.
2. Select "Show" softkey.
3. Select the "Enable Trace 2" menu item.
The R&S Spectrum Rider shows the second trace. The second trace is in a differ-
ent color. To show that the second trace is active, the R&S Spectrum Rider dis-
plays the current active trace "Trace 1" or "Trace 2" on the "Parameter View".
After you have activated the second trace, this trace becomes the active one. All
actions (like changing the detector or trace mathematics) apply to the active trace.
4. Select "Select Trace" softkey.
Trace 1 becomes the active trace.
You can put both traces into the internal memory of the R&S Spectrum Rider and
restore them later. Note that the memory trace 1 and memory trace 2 have a lighter
tone color.