Command Reference
R&S
®
ZNB/ZNBT
1287
User Manual 1173.9163.02 ─ 53
Example:
SYST:ERR:DISP:STATe ON; ERR ON; WARN:OFF; INFO
OFF
Display information popups for system errors, but not for warn-
ings and information messages.
Manual operation:
See
"Show Info Messages / Show Warning Messages / Show
SYSTem:ERRor:DISPlay:STATe
<Boolean>
Globally defines whether instrument events shall be indicated by information popups.
Display of popups can be limited to certain event types using commands
, and
Parameters:
<Boolean>
ON | OFF - globally enable or disable the display of information
popups
*RST:
ON
Example:
See
Manual operation:
See
SYSTem:FIRMware:UPDate
<VNASetupFile>
Installs a firmware version stored in a VNA setup file (
*.exe
) on the analyzer. The
installation is automatic and does not require any further action.
Setting parameters:
<VNASetupFile>
String variable for the name and directory of a VNA setup file.
Example:
SYST:FIRM:UPD 'C:
\Users\Public\Setup\ZNB_ZNBT_3.00.exe'
Install firmware version V3.00 from the setup file stored in the
public directory of the analyzer's internal hard disk.
Usage:
Setting only
SYSTem:FPReset
Performs a factory preset of all instrument settings (i.e. all open recall sets) or of the
active recall set, depending on the
SYSTem:PRESet:SCOPe
settings, and deletes
channel 1 in the active recall set. As a result, the active recall set contains no chan-
nels, traces, and diagram areas.
SCPI Command Reference