Reflection Measurements
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 2.4
E-1
Instrument Calibration
The analyzer provides a wide range of sophisticated calibration methods for all types of measurements.
Which calibration method is selected depends on the expected system errors, the accuracy
requirements of the measurement, on the test setup and on the types of calibration standards available.
In the following we assume that the calibration kit ZV-Z21 contains an appropriate male short standard
with known physical properties. With a single short standard, it is possible to perform a normalization,
compensating for a frequency-dependent attenuation and phase shift in the signal path.
Due to the analyzer's calibration wizard, calibration is a straightforward, menu-guided process.
1. Unscrew the DUT and connect the male short standard from calibration kit ZV-
Z21.
2. In the CHANNEL keypad, press CAL to open the calibration menu.
3. Activate
Start Cal – One Port P1 – Normalization (Short)
to open the calibration
wizard for the selected calibration type
.
4. In the first dialog of the wizard, select the calibration kit (here: ZV-Z21) and the
test port connector (here: N 50
Ω
(f), corresponding to a male calibration
standard), and click
Next
.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......