Reflection Measurements
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 2.2
E-1
Reflection Measurements
In a reflection measurement, the analyzer transmits a stimulus signal to the input port of the device
under test (DUT) and measures the reflected wave. A number of trace formats allow you to express and
display the results. depending on what you want to learn from the data. Only one analyzer test port is
required for reflection measurements.
In the following example, the analyzer is set up for a reflection measurement, a frequency sweep range
and measurement parameter is selected, the instrument is calibrated and the result is evaluated using
various formats.
Instrument Setup for Reflection Measurements
In order to prepare a reflection measurement, you have to connect your DUT (which is assumed to
have a male N 50
Ω
connector) to one of the (equivalent) analyzer test ports. Besides, it is
recommended to preset the instrument in order to set it to a definite, known state.
1. Proceed as described in section
Starting the Analyzer
and Shutting Down
to switch on the instrument and
start the NWA application.
2. Connect the input port of your DUT to test port 1 of
the network analyzer.
3. Press the preset key in the SYSTEM keypad to
perform a factory preset of the analyzer.
The analyzer is now set to its default state. The default
measured quantity is the transmission S-parameter S
21
. This
quantity is zero in the current test setup, so the trace shows
the noise level.
Press the TRACE SELECT key in the TRACE keypad and use the softkeys in the
Trace Select
menu if you wish to create a new trace or a new diagram area.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......