R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.191
E-6
Triggered Meas.
Sequence
Alternate On
Alternate Off
Sweep
Trigger event starts n partial sweeps over all
sweep points.
Trigger event starts m complete measurements at
consecutive sweep points.
Sweep Segment
Trigger event starts n partial sweeps over the
next segment.
Trigger event starts complete measurements at all
consecutive sweep points in the segment.
Point
All partial measurements of each sweep point
are carried out one after another.
All partial measurements of each sweep point are
carried out one after another.
Partial
Measurement
Each partial measurement is carried out for all
sweep points.
All partial measurements of each sweep point are
carried out one after another
Spurious Avoidance
The
Spurious Avoidance
submenu
defines whether the analyzer measures with a local oscillator
frequency LO below or above the RF input frequency.
This feature can be used to eliminate known spurious components in the input signal that can distort the
measurement, especially in the low frequency range.
•
In
Auto
mode, the analyzer auto-selects the local oscillator frequency, depending on the
receiver (RF) frequency and the test port. This mode systematically avoids known spurious
signals provided that no frequency conversion occurs in the test setup.
•
LO > RF
means that the LO frequency is always above the measured RF frequency. This mode
is appropriate for avoiding single, known spurious signals.
•
LO > RF
means that the LO frequency is always below the measured RF frequency. This mode
is appropriate for avoiding single, known spurious signals.
In the presence of several spurious signals, setting the
Spurious Avoidance
parameter globally
may not be sufficient. To improve the result, perform a
Segmented Frequency
sweep and assign
independent LO frequencies to the individual sweep segments.
Application example
Consider the following test setup with strongly reflecting DUT (e.g. a bandpass in its stop band) that is
measured in transmission. a
1
is generated at a frequency RF.The reflected wave b
1
falls into the
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......