
R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.149
E-6
Triggered Meas. Sequence
Selects the measurement cycle or sequence of actions to be triggered.
•
Sweep
means that each trigger event starts an entire sweep, according to the current sweep
configuration.
•
Sweep Segment
means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than
Segmented Sweep
is active, this option is equivalent
to
Sweep
.
•
Point
means that each trigger event starts the measurement at the next sweep point.
•
Partial Measurement
means that each trigger event starts the next partial measurement at the
current or at the next sweep point. If each measurement points requires only one partial
measurement , this option is equivalent to
Point.
Relation to other sweep settings
Some sweep settings are logically incompatible with a particular selection of the
Triggered Meas.
Sequence:
•
If a
Time
sweep is performed, the
Triggered Meas. Sequence
is always a sweep.
•
Alternate
sweep mode only makes sense if the triggered measurement sequence comprises
more than one sweep point. If
Point
or
Partial Measurement
is selected,
Alternate
sweep mode
is switched off and vice versa.
The period of the trigger events should be adjusted to the triggered measurement sequence. If
the analyzer receives a trigger event while the last sequence is still running, the trigger event is
skipped and a message Trigger event missed is displayed.
Remote control:
TRIGger[:SEQuence]:LINK
Trigger Delay
Specifies a delay time between the trigger event and the start of the measurement.
The Delay time entered must be zero or positive, indicating that the trigger event precedes the start of
the measurement (Posttrigger).
Remote control:
TRIGger[:SEQuence]:HOLDoff
Single (All Chans)
Toggles between single sweep and continuous sweep mode.
•
In continuous mode
(Single (All Chans)
not selected), the analyzer measures continuously,
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......