Show as
selects the physical unit of the displayed trace. It is possible to display the measured
Voltage
U or convert the wave quantity into an effective power according to P = V
2
/Re(Z
0
). Z
0
denotes the reference impedance of the source port (for wave quantities a
n
) or of the receive
port (for wave quantities b
n
). The reference impedances are defined in the
Port Configuration
dialog.
•
Source Port
selects one of the available test ports of the analyzer as a source of the stimulus
signal.
Remote
control:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "Pmtr1D1" |
...
Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "Pmtr1D1"
|...
Format
The
Format
submenu
defines how the measured data is presented in the graphical display.
•
dB Mag
selects a Cartesian diagram with a logarithmic scale of the vertical axis to display the
magnitude of a complex measured quantity.
•
Phase
selects a Cartesian diagram with a linear vertical axis to display the phase of a complex
measured quantity in the range between –180 degrees and +180 degrees.
•
Smith
selects a Smith diagram to display an S-parameter or ratio.
•
Polar
selects a polar diagram to display an S-parameter or ratio.
•
Delay
calculates the group delay from an S-parameter or ratio and displays it in a Cartesian
diagram.
•
Aperture
sets a delay aperture for the delay calculation.
•
SWR
calculates the Standing Wave Ratio from the measured reflection S-parameters and
displays it in a Cartesian diagram.
•
Lin Mag
selects a Cartesian diagram with a linear scale of the vertical axis to display the
magnitude of the measured quantity.
•
Real
selects a Cartesian diagram to display the real part of a complex measured quantity.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......