Calibration Overview
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 3.38
E-1
Calibration
Type
Standards Parameter
s
Error Terms
General
Accuracy
Application
TRL (2-port)
Reflect (at
both ports),
Through,
Line
All
See TOM
High, high
directivity
Reflection and
transmission
measurements on DUTs
with 2 ports, especially
for planar circuits.
Limited bandwidth.
TNA (2-port)
Through,
Attenuation,
Symmetric
network
All
See TOM
High, lowest
requirements on
standards
Reflection and
transmission
measurements on DUTs
with 2 ports, especially
for planar circuits.
Enhanced
TOM (2-port, 3-
port or 4-port)
Open,
Match (at
both ports),
Through
All Reflection
tracking,
Source match,
Directivity,
Load match,
Transmission
tracking
High
Reflection and
transmission
measurements on DUTs
with 2, 3, or 4 ports,
especially at high
frequencies.
Enhanced TSM
(2-port, 3-port
or 4-port)
Short,
Match (at
both ports),
Through
All Reflection
tracking,
Source match,
Directivity,
Load match,
Transmission
tracking
High Reflection
and
transmission
measurements on DUTs
with 2, 3, or 4 ports,
especially at high
frequencies.
1) Or any other 3 known one-port standards. To be used in a guided calibration, the known standards must be declared to be
open, short, and match irrespective of their properties.
2) Or any other known two-port standard. See remark above.
The calibration type must be selected in accordance with the test setup. Select the calibration
type for which you can obtain or design the most accurate standards and for which you can measure
the required parameters with best accuracy.
Normalization
A normalization is the simplest calibration type since it requires the measurement of only one standard
for each calibrated S-parameter:
•
One-port (reflection) S-parameters (S
11
,
S
22,
...) are calibrated with an open or a short standard
providing the
reflection tracking
error term.
•
Two-port (transmission) S-parameters (S
12
,
S
21,
...) are calibrated with a through standard
providing the
transmission tracking
error term.
Normalization means that the measured S-parameter at each sweep point is divided by the
corresponding S-parameter of the standard. A normalization eliminates the frequency-dependent
attenuation and phase shift in the measurement path (reflection or transmission tracking error). It does
not compensate for directivity or mismatch errors. This limits the accuracy of a normalization.
Full One-Port Calibration
A full one-port calibration requires a short, an open and a match standard to be connected to a single
test port. The three standard measurements are used to derive all three reflection error terms:
•
The short and open standards are used to derive the
source match
and the
reflection
tracking
error terms.
•
The match standard is used to derive the
directivity
error.
A full one-port calibration is more accurate than a normalization but is only applicable for reflection
measurements.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......