R&S
®
ZVA/ZVB/ZVT
Calibration Overview
1145.1084.12 3.37
E-1
The analyzer provides a large number of predefined cal kits but can also import cal kit files and create
new kits:
•
A selection of predefined kits is available for all connector types. The parameters of these kits are
displayed in the
Add/Modify Standards
dialog, however, it is not possible to change or delete the
kits.
•
Imported and user-defined kits can be changed in the
Calibration Kits
dialog and its various sub-
dialogs.
Calibration kits and connector types are global resources; the parameters are stored independently and
are available irrespective of the current setup.
Calibration Types
The analyzer provides a wide range of calibration types for one, two or more ports. The calibration
types differ in the number and types of standards used, the error terms, i.e. the type of systematic
errors corrected and the general accuracy. The following table gives an overview.
Calibration
Type
Standards Parameter
s
Error Terms
General
Accuracy
Application
Reflection
Normalization
Open or
Short
S
11
(or S
22
, ...)
Reflection tracking
Low to medium
Reflection
measurements on any
port.
Transmission
Normalization
Through S
12
, S
21
(or S
13
,...)
Transmission
tracking
Medium Transmission
measurements in any
direction and between
any combination of
ports.
Full One-Port
Open,
Short,
Match
1)
S
11
(or S
22
, ...)
Reflection tracking,
Source match
Directivity,
High Reflection
measurements on any
port.
One-Path Two-
Port
Open,
Short,
Match
1)
(at
source
port),
Through
2)
S
11
, S
21
(or S
22
,...)
Reflection tracking,
Source match,
Directivity,
Transmission
tracking
Medium to high
Unidirectional
transmission
measurements in any
direction and between
any combination of
ports.
TOSM
(2-port, 3-port
or 4-port)
Open,
Short,
Match
1)
(at
each port),
Through
2)
(between all
combination
s of 2 ports)
All Reflection
tracking,
Source match,
Directivity,
Load match,
Transmission
tracking
High Reflection
and
transmission
measurements on DUTs
with 2, 3 or 4 ports;
classical 12-term error
correction model.
TOM (2-port)
Open,
Match (at
both ports),
Through
All Reflection
tracking,
Source match,
Directivity,
Load match,
Transmission
tracking
High, implicit
verification
Reflection and
transmission
measurements on DUTs
with 2 ports.
TRM (2-port)
Reflect
(equal at
both ports),
Match,
Through
All
See TOM
High
Reflection and
transmission
measurements on DUTs
with 2 ports, especially
in test fixtures.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......