Common Commands FSE
1065.6016.12
6.4
E-16
Common Commands
The common commands are taken from the IEEE 488.2 (IEC 625-2) standard. Same commands have
the same effect on different devices. The headers of these commands consist of an asterisk "*" followed
by three letters. Many common commands refer to the status reporting system which is described in
detail in Chapter 5.
Command
Designation
Parameter
Remark
*CAL?
Calibration Query
query only
*CLS
Clear Status
no query
*ESE
Event Status Enable
0 to 255
*ESR?
Standard Event Status Query
0 to 255
query only
*IDN?
Identification Query
<string>
query only
*IST?
Individual Status Query
0 to 255
query only
*OPC
Operation Complete
*OPT?
Option Identification Query
query only
*PCB
Pass Control Back
0 to 30
no query
*PRE
Parallel Poll Register Enable
0 to 255
*PSC
Power On Status Clear
0 | 1
*RST
Reset
no query
*SRE
Service Request Enable
0 to 255
*STB?
Status Byte Query
query only
*TRG
Trigger
no query
*TST?
Self Test Query
query only
*WAI
Wait to continue
no query
*CAL?
CALIBRATION QUERY triggers a calibration of the instrument and subsequently query the
calibration status. Any responses > 0 indicate errors.
*CLS
CLEAR STATUS sets the status byte (STB), the standard event register (ESR) and the EVENt-part
of the QUEStionable and the OPERation register to zero. The command does not alter the mask and
transition parts of the registers. It clears the output buffer.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......