FSE SENSe Subsystem
1065.6016.12
6.167
E-16
SENSe:BANDwidth Subsystem
This subsystem controls the setting of the instrument’s filter bandwidths. Both groups of commands
(BANDwidth and BWIDth) perform the same functions.
COMMAND
PARAMETERS
UNIT
COMMENT
[SENSe<1|2>]
:BANDwidth | :BWIDth
[:RESolution]
:AUTO
:MODE
:FFT
:RATio
:VIDeo
:AUTO
:RATio
:DEMod
:PLL
<numeric_value>
<Boolean>
ANALog|DIGital
<Boolean>
<numeric_value>
<numeric_value>
<Boolean>
<numeric_value> | SINe | PULSe |
NOISe
<numeric_value>
AUTO | HIGH | MEDium | LOW
HZ
--
--
--
--
HZ
--
--
HZ
option FFT Filter
Vector Signal Analysis
:[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution] <numeric_value>
This command defines the analyzer’s resolution bandwidth.
Example:
":BAND 1MHz"
Features:
*RST value:
- (AUTO is set to ON)
SCPI:
conforming
Modes:
A, VA, BTS, MS
The values for the resolution bandwidth are rounded in 1 | 2 | 3 | 5 steps.
In the GSM BTS/MS ANALYZER mode with option FSE-K11/K10, the command is available for
POWER vs. TIME measurement. In this case, the parameters DEFault (bandwidth setting according
to GSM standard), 300KHZ and 1MHZ are permitted.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......