Roche Diagnostics
C-40
Operator’s Manual · Version 1.0
6 Special operations
URISYS 2400
Calibration
Specific gravity (SG) calibration
The specific gravity of the sample changes with temperature. Since the
URISYS 2400 analyzer measures the temperature of the flow cell, there may be
a difference in temperature between the sample and the flow cell. Therefore,
the temperature of the urine sample and the calibration solutions must always
be at room temperature.
a
To prepare the calibration solutions
Use deionized water and reagent-grade NaCl to prepare the following
solutions, and place them in a rack:
Set the rack on the input line.
a
To calibrate the SG meter
1
Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
2
Press <Tools>. The [Tools] screen is displayed.
3
Press <User Calibration>. The [User Calibration] screen is displayed.
4
Press <SG Calibration>. The [SG Calibration] screen is displayed.
5
Press <Calibrate>. The [SG Calibration] window appears.
6
Press <Execute>.
Sample analysis can start even if the SG calibration is unsuccessful. If SG
calibration is unsuccessful, the previous calibration values are used.
Positions 1 and 3
Deionized water
Positions 2 and 4
NaCl solution 10.0 g/dL
Position 5
NaCl solution 5.0 g/dL
After the calibration process is completed, “Calibration OK” or “Calibr. Failed” is
displayed on the [SG Calibration] screen. If “Calibr. Failed” is displayed, repeat the
SG calibration.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......