5 Daily operation
URISYS 2400
Analysis and documentation of results
Edit result data
a
To edit routine and STAT sample data
1
Search for the requested sample by using the sequence or ID number, or
the <Prev>/<Next> buttons.
2
Press <Edit Data>. The [Edit Data] screen is displayed.
3
Press “<” or “>” to select the test parameter to be edited.
4
Press “+” or “-” to increase or decrease the result.
5
Press <Save> when editing is completed. The [Confirmation] window
appears.
6
Press <Yes> to return to the [Edit Data] screen. Check if the changes are
correct.
7
Press <PgUp> to return to the [Routine Data Monitor] screen.
When result data are edited, they are printed with an “!” edit mark.
Print routine and STAT
sample data
This print option is available for batch printing. For more information on
printing in real time, see:
a
To print routine and STAT sample data
1
Press <Type> on the [Routine Data Monitor] screen to select the type of
data to be displayed (routine or STAT).
2
Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3
Enter the range of sequence numbers to be printed, using numbers
between 1 and 1000 for routine sample data, or between 1 and 200 for
STAT sample data. Enter 0 to print all sample data.
4
Press <Print>. The [Confirmation] window appears.
5
Press <Data Select> to select the category of results to be printed (All,
Abnormal, Sieve, Normal).
6
Press <OK> to start printing.
o
To edit sample data, the analyzer must be in Stand-By status
o
Data cannot be edited if a sample is short or empty, or if a test strip error
occurred, (see Table E-1 on page E-2 and “Instrument alarms
(messages)” on page E-12).
“To select real time printouts and specify the result data to be printed” on page C-7
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......