Roche Diagnostics
Operator’s Manual · Version 1.0
B-21
URISYS 2400
3 Software
Installation
Range table
The operator can view the concentration ranges, reflectance values,
wavelengths and compensation data for each test parameter. The operator
gains access to the range and reflectance values by using the password that is
supplied with the analyzer.
The wavelength data (W) that are displayed are fixed values. These values
cannot be changed.
The compensation data (C) designate factors for the compensation of the
intrinsic urine color, for example, 0 means no color compensations. Higher
numbers mean stronger compensation.
Changing ranges leads to result reports which differ from the default
concentration ranges as shown in:
The concentration ranges and reflectance values can be reset to the default
values using the <Default> button.
Figure B-11 Screen showing Range Table settings and options
Changing reflectance values leads to different evaluation sensitivities of the
respective test parameters. Lowering the reflectance value of the negative (NEG)
range leads to a decrease of the sensitivity of the test evaluation, and vice versa. In
this way, the sensitivity can be adjusted to the requirements of the individual
laboratory.
The correctness of results obtained after the user has altered the ranges or
reflectance values is not warranted by Roche Diagnostics. The user is responsible
for validating the consistency of results after changes have been made.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......