3 Software
URISYS 2400
Installation
a
To set the criteria for defining sieve and abnormal values
1
Press <Sieve & Abnormal> on the [Test Parameters] screen. The [Sieve &
Abnormal] screen is displayed listing the sieve criteria for each test.
2
Press <Select> to change the display of the items from sieve to abnormal
criteria and back.
3
Pr
ess <Set> to change the criteria. The [Password] window appears.
4
Enter the 4-digit password using the keyboard on the screen.
5
Press <Enter>. The [Sieve & Abnormal Setting] screen is displayed with
both the sieve and abnormal values listed for each test parameter,
beginning with specific gravity (SG).
6
Press <Test> to select another test parameter.
7
Press “<” or “>” to select sieve or abnormal settings.
8
Press <+> or <-> to select the desired concentration range.
9
Pre
ss <Enter> after each selection to accept the chosen values.
10
Press <PgUp> to return to the [Sieve & Abnormal] screen and check if the
settings correspond to your requirements.
11
Press <PgUp> to return to the [Test Parameters] screen.
Print order
The operator can define the order of the test parameters on the screen and on
the result printout, in order from 1 to 12. If an item is not to be printed, enter
zero.
Figure B-10 Screen showing Print Order settings and options
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......