Roche Diagnostics
B-10
Operator’s Manual · Version 1.0
2 Hardware
URISYS 2400
Installation
Environmental conditions
The URISYS 2400 analyzer must be used in an environment that meets the
following conditions:
o
Free from excessive dust
o
Well-ventilated area
o
Not exposed to direct sunlight
o
Floor that is level (with an inclination of less than 1.5 degrees in every
direction) and firm
o
Table that can support the mass of the instrument, which is approximately
85 kg (187 lbs)
o
Ambient temperature ranging from 15-30
°
C (59-86
°
F), with an acceptable
variation of
±
2
°
C (
±
3
°
F) during measurement
o
Relative humidity ranging from 20% to 80%, without moisture
condensation
o
Free from vibrations
o
Availability of a 3-pin power supply socket within 3 m (9.8 ft) of the
analyzer
o
Free from abrupt mains voltage fluctuation (
±
10 V)
o
Well-distanced from a machine generating a high frequency voltage (for
example, a centrifuge)
o
Free from electromagnetic wave interference
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......