Roche Diagnostics
B-6
Operator’s Manual · Version 1.0
2 Hardware
URISYS 2400
Installation
Power supply
A standard 230 V/50 Hz or 115 V/60 Hz power line is required to operate the
analyzer. It must be connected to a grounded power supply socket (protection
class 1).
Although the maximum power consumption of the instrument is 200 VA, the
external power supply should fulfill the following minimum requirements to
cover for peaks at switching on.
Electric power supply capacity:
o
Minimum 500 VA
o
Minimum 5 Amp, 3-pin power supply socket
The power switch and power supply socket are located on the right side of the
analyzer.
Water supply
An external water supply tank for deionized water is supplied with the
URISYS 2400. The deionized water must be germ free and have an electric
conductance value that is less than or equal to 1 µS/cm.
Figure B-2 Power switch and power supply socket
A
Power switch
B
Power supply socket
A
B
Do not connect the analyzer directly to a pressurized water supply.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......