Roche Diagnostics
Operator’s Manual · Version 1.0
E-9
URISYS 2400
8 Data alarms (flags)
Reset the analyzer
Reset the analyzer
The reset procedure is used as a troubleshooting tool.
a
To reset the analyzer
1
Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
2
Press <Maintenance>. The [Maintenance] screen is displayed.
3
Press <Reset>. The [Reset] window appears.
4
Press <Execute> to start the resetting.
During Reset, the analyzer clears unprocessed racks from the cross-feed line,
discards test strips still present on the TS transfer base, performs a rinsing
cycle of the liquid handling system and resets these modules to their home
position.
Start
If the analyzer does not switch on at all, check that:
o
The main power is on.
o
The power cable is connected properly.
Contact Technical Support if the analyzer still does not respond.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......