Roche Diagnostics
Operator’s Manual · Version 1.0
D-7
URISYS 2400
7 General maintenance actions
Maintenance menu
To prevent overflow of the liquid waste tank, it is recommended to install an
overflow sensor.
Clean the external parts of the analyzer
Use a cloth dampened with water or with a liquid cleaning agent to clean the
rack input line, the rack trays, and the outer surface of the analyzer. Apply a
disinfectant if necessary.
For cleaning the touch screen, switch off the analyzer. Wipe the touch screen
with a dry cloth.
If contamination is heavy, clean with a cloth moistened with a neutral
detergent.
Air purge
The air purge function is used to remove air bubbles from the syringes, flow
cell, internal tubing, and sample needle. During first time installation, this
function is used to fill the water supply tube and the liquid handling system
with deionized water.
In addition, use the air purge function after refilling the water supply tank and
if air bubbles are present in the syringes.
a
To remove air bubbles
1
Set the analyzer to Stand-By mode.
2
Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
3
Press <Maintenance>. The [Maintenance] screen is displayed.
4
Press <Air purge>. The analyzer performs several rinsing cycles.
5
Press <Execute>.
Do not use volatile solvents such as benzene. Do not scratch the surface of the
touch screen.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......