7 General maintenance actions
Maintenance procedures
Every week
The following are maintenance procedures to be performed weekly:
o
Check the water supply filter for growth of bacteria or algae.
o
Clean the rack input line.
o
Clean the rack trays.
Every month
The following are maintenance procedures to be performed monthly:
o
Clean the outer parts (housing) of the analyzer.
o
Clean or change the water supply filter.
Every three months
The following is the maintenance procedure to be performed every three
months:
o
Change the seals on both syringes.
Unplug the analyzer when changing the seals on the syringes.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......