4 Measurement
4-6
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
4.2.1 Syringe mode
Improper heparinization of syringes with liquid heparin may cause false results.
ISE parameters are particularly susceptible.
Please contact Roche customer support regarding this matter.
•
Open the flap to the designated syringe position
(half-opened position). The aspiration process is started.
Fig. 2
•
Attach the syringe (always remove the cannula first).
Be sure that the needle sufficiently penetrates the interior of the syringe (or ampoule)
in order to remove the sample without air bubbles.
Be sure that the needle does not make contact with the fill port while inserting the nee-
dle into the interior of the syringe.
If the opening of the needle is covered by the fill port, the aspiration is blocked and the
sample is rejected!
If using an ampoule instead of a syringe (during a QC measurement), be sure that the
ampoule does not make contact with the fill port when inserting the needle into the
ampoule. The sharp edge can cause damage to the fill port.
The clot catcher cannot be used in "Syringe mode!"
Fig. 3
•
The sample is automatically aspirated (it is unnecessary to press any buttons).
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...