1 Introduction
1-6
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
1.4
Safety instructions for specific dangers
1.4.1 Handling samples
Respect the necessary hygienic regulations when handling samples because samples may
contain dangerous viruses.
For more detailed information, please see chapter 4, "
Measurement
."
1.4.2 Disposal of waste water, bottles, electrodes, and the instrument
Dispose waste water, bottles, electrodes and the instrument according to local and/
or labour regulations (biologically contaminated—hazardous waste!)
1.4.3 Decontamination
The purpose of this procedure is to minimize risk when handling items that were in contact
with biological samples.
Roche recommends following a decontamination procedure in addition to regulations spe-
cific to the laboratory.
These decontamination procedures should be performed periodically to minimize the risk
of infections (incl. hepatitis virus and HIV).
Always wear gloves!
For more detailed information about decontamination, please see chapter 6, "
Mainte-
nance."
1.5
Handling solutions
Store the OMNI C solutions according to the specified packaging requirements. The solu-
tions should be adapted to the ambient temperature before use.
The shelf life of the solutions is limited.
Please read the bottle label and the packaging for the correct storage temperature and the
maximum shelf life.
CAUTION! DO NOT FREEZE!
If frozen, the solution's concentration may change and cause calibration errors!
Do not use damaged C3 fluid packs! Do not mix the individual components!
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...