7 Trouble shooting
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
7-9
7.2.19 3024 Flash memory full
Cause
•
The internal flash memory has less than 8 KB space left for saving additional data
Remedy
•
Delete data records (database entries, protocols, log data) in order to free up additional
memory (see chapter 8, "Operating modes", section "Database" or Reference Manual,
chapter 3, "Operating modes", section "Setup – Displays and reports")
Important! In order to effectively free up additional memory, the functions "Delete dat"
and "Optimize database" have to be activated in this order!
7.2.20 3025 PCMCIA memory full
Cause
•
The PCMCIA card has less than 8 KB space left for saving additional data
Remedy
•
Delete data records from the PCMCIA card in order to free up additional memory
capacity
- insert the card into a PC that has a suitable port, export the data, and then delete the
data from the card
- reformat the card in the Roche OMNI C (see Reference Manual, chapter 3, "Operating
modes", section "System – Test – PC Components – PCMCIA card)
7.2.21 3026 Data object error
Cause
•
The data access onto objects in the analyzer area failed, the the correct operation of the
Roche OMNI C can not be guaranteed
Remedy
•
Press the "Reboot" button
•
In case of recurrence, call customer service (the electronics is defective, change compo-
nents if necessary)!
7.2.22 3028 Hardware error
Cause
•
Electronic components do not respond properly
Remedy
•
Wait! These errors are automatically self-repaired!
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...