7 Trouble shooting
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
7-5
7.2.6 3006 Temperature error
Cause
•
The module temperature is outside of the specified range:
•
Measuring chamber (left and right): 37.00 °C ± 0.2 °C
•
Measuring chamber cover: 37.00 °C ± 0.2 °C
•
tHb-/
S
O
2
module: 37.00 °C ± 0.2 °C
•
A heating device is defective
•
The measuring chamber cover cable is defective
•
A temperature sensor is defective
Remedy
•
Reduce / raise the room temperature
•
In case of recurrence, call customer service!
7.2.7 3009 Conductivity cal. error
Cause
•
The conductivity calibration has failed
Remedy
•
Press the "OK" button (start a system calibration)
•
In case of recurrence, call customer service!
Consequence actions
•
System calibration
7.2.8 3010 AQC cover open
Cause
•
The AutoQC cover is open
•
The AutoQC cover sensor (Hall sensor) is defective
Remedy
•
Close the AutoQC cover
•
Check the AutoQC cover sensor ("More functions > System > Test > Control sensors >
Monitoring Sensors").
•
In case of recurrence, call customer service (change components if necessary)!
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...