7 Trouble shooting
7-4
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
7.2.3 3003 Flap open
Cause
•
The flap was opened:
- during a measurement
- during a calibration
- during another system stop
- in menu "System" and menu "System" is being closed
•
The flap detection is defective
Remedy
•
Close flap
•
Check function of the flap detection board ("More functions > System > Test > Control
sensors > Monitoring sensors"), change if necessary
•
In case of recurrence, call customer service!
Consequence actions
•
Wash
7.2.4 3004 Analyzer error
Cause
•
The measurement progress was incorrect
Remedy
•
Press "OK" button
•
Switch the analyzer off/on
•
In case of recurrence, call customer service (the electronics is defective, change compo-
nents if necessary)!
Consequence actions
•
Wash
7.2.5 3005 Memory error
Cause
•
Fundamental software functions can not be performed (memory problems, file system
problems), the correct operation of the Roche OMNI C can not be guaranteed
Remedy
•
Press the "Reboot" button
•
In case of recurrence, call customer service (the electronics is defective, change compo-
nents if necessary)!
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...