General Overview
15
A
A
2.3.2
Maintenance Items
Note:
䉴
To ensure that they work correctly, you must grease the O-rings once per week (or
after every 20 purification runs). If the O-rings malfunction, the instrument nozzles
may leak. For details, please see the Maintenance section in Chapter C.
䉴
Perform a leakage test every two months (or after every 50 purification runs) to deter-
mine whether the O-rings should be changed. For details please see the Maintenance
section in Chapter C.
There is an O-Ring Exchange Tool available, which simplifies the O-ring changing pro-
cess. Please contact your local Roche Diagnostics representative if you need O-rings
changed or want one of these tools.
䉴
Elution Tubes (35 tubes per package, 2.0 ml each, with stand, with barcode)
䉴
Caps for Elution Tubes (35 caps per caps)
䉴
Tip Trays (including Piercing Tool and tips). The tips are placed in the following order
into the Tip Tray: one Piercing Tool, one small tip, two large tips.
Cat. No.
Purification Kit
Contents
03 561 429 001
MagNA Pure LC O-Ring Maintenance Kit
O-rings (12 x 8),
High Vacuum Grease (1)
03 753 166 001
MagNA Pure Compact Tip Tray Kit
10 Tip Trays
Kit content:
Installation of the Instrument
Additional Required Items for Operation and Maintenance
MagNA Pure Compact Manual.book Seite 15 Dienstag, 28. Februar 2006 4:16 16
Summary of Contents for MagNA Pure Compact Instrument
Page 2: ......
Page 6: ......
Page 11: ......
Page 36: ......
Page 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Page 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Page 129: ......