Now you can remove both the Liquid Waste
Tank and the Drip Tray.
To remove and place the Liquid Drop
Catcher on the magnet unit, first remove the
Cartridge Rack:
The area around the instrument should be periodically checked to ensure that air flow
around the instrument is unrestricted and that books, papers or other items do not inter-
fere with air flow.
If a sample or other biohazardous materials is spilled on the system or any of the system
racks, the area or racks should be cleaned with a 10% (v/v) bleach solution (0.5% sodium
hypochlorite) followed by thorough cleaning with 70% ethanol.
Commercial liquid household bleach typically contains sodium hypochlorite at a concen-
tration of 5.25%. A 1:10 dilution of household bleach will produce a 0.5% sodium
hypochlorite solution.
After using bleach, ventilate the instrument for at least 1 hour.
Never clean the cartridge rack in a diswasher. All other removeable accessories are
dishwasher proof. Always use only detergents that are compatible with the instrument’s
surface. If in doubt, please contact your local Roche representative!
Cleaning Instructions and UV Decontamination
Cleaning
MagNA Pure Compact Manual.book Seite 84 Dienstag, 28. Februar 2006 4:16 16
Summary of Contents for MagNA Pure Compact Instrument
Page 2: ......
Page 6: ......
Page 11: ......
Page 36: ......
Page 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Page 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Page 129: ......