6
• Control Testing and Quality Control
98
6.2
Performing a liquid quality control test
1
Place the meter on a level, vibration-free surface or
hold it in your hand so it is roughly horizontal.
2
Power the meter on by pressing the
button for
approximately 1 second.
You can also power on the meter directly by inserting a
test strip or connecting the power adapter.
3
Wait until the Main Menu is displayed, or log on as
4
Check the battery level.
5
Check that the date and time are correct. Correct
any wrong entries as described in Chapter 4,
■
If the battery icon turns red (one bar left), there may
not be enough power left for another test.
■
If there are no bars left in the battery icon, you cannot
perform any more tests. Power the meter off using the
button.
In both cases, restore power by inserting new batteries,
recharging the battery pack, or using the power adapter.
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......