5
• Testing a Blood Sample
90
The meter performs an automatic quality control test on
the test strip before it displays the test result. “QC”
appears in the display.
Following a successful outcome of the quality control test,
a checkmark appears after “QC.”
The result is displayed in the unit you chose when setting
up the meter. It is automatically saved.
The valid measuring ranges for CoaguChek XS PT test
strips are:
■
INR: 0.8–8.0
■
%Q: 120–5
■
Sec: 9.6–96
Results that are above or below the measuring range are
indicated by the symbols > (above) or < (below).
When interpreting results, refer to the detailed informa-
tion on limitations and interferences included in the limi-
tations section of the test strip package insert.
If a “c” is displayed along with the result:
This may occur if the hematocrit value is very low or due
to erroneous blood collection (e.g., wet hands). Repeat
the measurement and make sure that the patient's
hands are dry. If the message persists, perform a
hematocrit check.
Test
09:15 am
04/19/2012
P-ID: PID01
Code: 184
Smith, Henry
Op:
Operator 1
Par:
PT
QC
Test
09:15 am
04/19/2012
P-ID: PID01
Code: 184
2.5
INR
Smith, Henry
Op:
Operator 1
Par:
PT
21
%Q
04/19/2012 09:14 am
Test
09:15 am
04/19/2012
P-ID: PID01
Code: 184
Smith, Henry
Op:
Operator 1
Par:
PT
04/19/2012 09:14 am
> 8.0
INR
Test
09:15 am
04/19/2012
P-ID: PID05
Code: 208
1.6
INR
Müller, David
Op:
Schulz
Par:
PT
Asymptomatic
Doctor Noti
fi
ed
NEW MEDICATION
04/19/2012 09:14 am
C
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......