Using the OTS function
A request for an Observed Test Sequence comes from the
DMS. The presence of the
icon on the
Patient Test
button indicates a pending OTS request.
Observer:
1
Touch
Patient Test
.
In the
Pat. Test -OTS
menu, the
Patient Test
button is
grayed out (disabled) until the observer has logged in.
2
Touch
Observer Login
.
3
Wait until the observer list is displayed.
4
Select your observer ID by touching the corre-
sponding button, or scan your operator ID (which is
also your observer ID in this case).
5
Enter the (optional) password.
6
After you enter your password, touch
to log in.
The
Pat. Test - OTS
menu is displayed again. The
Patient Test
button is now active.
7
Hand the meter to the operator who can now per-
form the patient test under supervision.
Observer
09:15 am
04/19/2012
Ann Casey
Alan Smithee
Scan
Pat. Test - OTS
09:15 am
04/19/2012
Observer Login
Patient Test
OTS enabled
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Logout
Pat. Test - OTS
09:15 am
04/19/2012
Observer Login
Patient Test
OTS enabled
Only operators with OTS observer rights are listed in the
Observer Login list.
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......