Disinfecting the meter
NOTE: Clean the meter prior to each
disinfection step
Remember to keep the meter on a level
surface prior to disinfecting and powered off.
Squeeze off excess disinfecting solution
or blot on a dry paper towel to remove
any excess disinfecting solution before
disinfecting the surface of the meter.
1
Use a fresh Super Sani-Cloth Germicidal
Disposable Wipe to disinfect by gently
wiping the outside of the meter three
times horizontally and three times
vertically and carefully wipe around the
test strip guide area, making sure that no
liquid enters the test strip guide.
2
Allow the surface of the meter to remain
damp with the recommended disinfecting
solution. The recommended contact time
is 2 minutes for the Super Sani-Cloth
Germicidal Disposable Wipe.
Cleaning and Disinfecting the Meter and the Lancing Device
CoaguChek Vantus User’s Manual, version 1.0