Automatic shutdown
If the battery pack is critically low, the meter will
shut down automatically.
r
To reset the meter
A meter reset should only be performed if all other
remedies have failed.
1
Place the meter on a level surface.
2
Press the ON/OFF button for at least 12 seconds
and then release it. After releasing the button, the
meter will start up again.
•
The meter powers off and on again.
•
The Roche logo is displayed. If the Roche logo
does not appear within 60 seconds, place the
meter in the Handheld Base Unit for a
minimum of 15 minutes to recharge the
battery.
•
The meter performs a system check.
•
The screens for entering the date and the time
appear.
3
Enter the date and time. After you have entered
the correct information, confirm each screen with
.
•
Unless all lockouts are disabled, the meter will
now be in QC Lockout due to the manually
entered date/time.
9 Troubleshooting
• 9
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......