Operating conditions
To ensure that the meter functions properly, please
observe the following guidelines:
■
Only use the meter at a room temperature between
12 °C and 32 °C (54 °F and 90 °F).
■
Only use the meter at a relative humidity between
10% and 85% (no condensation).
■
When operating the meter using the power adapter,
use only a voltage of 100 V to 240 V (± 10%) , 50/
60 Hz.
■
When testing, place the meter on a level, vibration-
free surface, or hold it so it is roughly horizontal.
Quality control
The meter has a number of built-in quality-control
functions:
■
A check of the electronic components and func-
tions every time the meter is powered on.
■
A check of the test strip temperature while a test is
in progress.
■
A check of the expiration date and lot information
on the test strip based on the code chip data.
■
An onboard quality control within every single test
strip.
Roche Diagnostics offers liquid quality controls for the
CoaguChek Pro II system. These controls are designed to
assist you in meeting regulatory compliance require-
ments at your facility.
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......