Maintenance and Care •
8
151
8.3
Recommended cleaning/disinfecting agents
Use only the following solutions for cleaning/disinfecting
the meter (housing and test strip guide).
■
A soft cloth slightly dampened (not wet) with a
small amount of liquid soap diluted in water
■
Rubbing alcohol (70% ethanol or isopropyl alcohol)
■
Alcohol-based disinfectant (a mixture of
1-propanol (400mg/g), 2-propanol (200 mg/g) and
glutaraldehyde (1.0 mg/g))
1
■
10% sodium hypochlorite solution (1 part bleach
to 9 parts de-ionized water, made fresh every
24 hours)
■
Disposable wipes containing quaternary ammo-
nium compounds up to 0.5 % (single compound or
mixture) in isopropyl alcohol (isopropanol) up to
55 %
1.
Sold in some countries under the name “Bacillol Plus.”
NOTICE
Do not use any other disinfectants/cleaning solutions on
the meter (housing or test strip guide). Use of other dis-
infectants/cleaning solutions could result in damage to
the meter.
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......