3 Hardware
Cedex Bio System
Hardware overview
The following table lists a few typical tubes that are suitable, and it gives the dead
volume for each of them.
Roche recommends using sample cups because they need much lower dead volume
compared to primary tubes. The following table lists recommended cups.
Cuvettes
WARNING
Incorrect results due to inappropriate tube and cup placement
Inappropriate tube and cup placement may lead to inaccurate pipetting and consequently
to incorrect results.
Make sure that the primary tubes are placed centrally and perfectly vertically in the holders
in the sample area and that they are inserted firmly.
Make sure that secondary tubes are placed centrally on the primary tubes and that they
rest fully on them.
NOTICE
Probe damage due to not removing primary tube caps
The probe is not designed to pierce tube caps. It can get damaged when trying to pierce
tube caps.
Always remove the caps of primary tubes before placing them on the instrument.
Tube name
Dead volume
13 x 75 mm
500
μ
L
13 x 100 mm
500
μ
L
16 x 75 mm
700
μ
L
16 x 100 mm
700
μ
L
Table A-6
Typical examples of suitable tubes
Cup name
Dead volume
Sample volume
Placement
1.5 mL reaction tube
100
μ
L
1.5 mL
Using adapter cups
Roche Diagnostics
Standard false bottom
tube
75
μ
L
2.5 mL
Directly on sample area
Table A-7
Typical examples of suitable tubes
All optical measurements are made using the same transparent
plastic containers, called cuvettes. Samples are automatically
transferred from a sample tube to the cuvettes on the cuvette ring.
Cuvettes are disposable to eliminate carry-over.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...