14 ISE maintenance
ISE maintenance actions
Replace ISE unit tubing
f
Utilities
>
Maintenance
>
Replace ISE Unit Tubing
Due to the abrasion by the ISE valve action, the ISE tubing needs to be replaced
periodically. Replacing the ISE tubing ensures the proper functioning of the ISE unit.
During the
Replace ISE Unit Tubing
maintenance action, the following
maintenance actions are automatically performed:
o
Prime ISE Reference and Calib.
o
Initialize ISE Reference Sensor
o
Condition ISE Tubing
o
Initialize ISE Unit
Operator time
Approximately 15 minutes.
Prerequisites
The system must be in
Standby
status.
Tools and materials required
m
Activator
m
Roche approved ISE unit tubing set
m
Tissues
a
To remove the ISE unit tubing
1
Select the maintenance action
Replace ISE Unit Tubing
.
2
Press .
The maintenance definition screen is displayed.
3
Press
to start the action.
The placement screen is displayed.
4
Place the Activator on the sample area position indicated on the screen.
5
Press
to confirm the placement.
The tubing is emptied.
Your are asked to open the covers.
6
Open the main cover, then the left service flap and the ISE cover.
Your are asked to remove the tubing adapters from the ISE solution bottles.
Always replace the ISE tubing with the help of the
Replace ISE Unit Tubing
maintenance
action.
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through working solutions
o
Infection by biohazardous materials
o
o
Personal injury or damage to the analyzer due to contact with instrument mechanism
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...