14 ISE maintenance
Cedex Bio System
ISE maintenance actions
Clean ISE tower automatically
f
Utilities
>
Maintenance
>
Clean ISE Tower Automatically
The ISE tower must be cleaned regularly to ensure the proper functioning of the ISE
unit. Cleaning includes deproteinizing and activating the ISE tower.
Operator time
Approximately 3 minutes.
Prerequisites
The system must be in
Standby
status.
Tools and materials required
m
Activator
m
ISE Deproteinizer
a
To clean the ISE tower automatically
1
Select the maintenance action
Clean ISE Tower Automatically
.
2
Press .
The maintenance definition screen is displayed.
3
Press
to start the action.
A placement list is displayed.
4
Place the Activator and the ISE Deproteinizer on the sample area positions
indicated on the screen.
5
Press
to confirm the placement and to start the action.
A message will inform you when the maintenance action is complete.
6
Remove the Activator and ISE Deproteinizer from the sample area.
7
Press .
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through working solutions
o
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...