Roche Diagnostics
E-46
Operator’s Manual · Version 3.1
14 ISE maintenance
Cedex Bio System
ISE maintenance actions
Condition ISE tubing
f
Utilities
>
Maintenance
>
Condition ISE Tubing
Conditioning the tubing with Activator ensures the proper flow of fluids.
Operator time
Approximately 3 minutes.
Prerequisites
The system must be in
Standby
status.
Tools and materials required
m
Activator
a
To condition the ISE tubing
1
Select the maintenance action
Condition ISE Tubing
.
2
Press .
The maintenance definition screen is displayed.
3
Press
to start the action.
A placement list is displayed.
4
Place the Activator on the sample area position indicated on the screen.
5
Press
to confirm the placement and to start the action.
A message will inform you when the maintenance action is complete.
6
Remove the Activator from the sample area.
7
Press .
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through working solutions
o
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...