Roche Diagnostics
Operator’s Manual · Version 3.1
A-29
Cedex Bio System
2 Introduction to the instrument
Daily operation
Forecast period
The forecast period is a configurable period of time. Calibrations that fall due within
this period will be performed collectively.
e
Typically you would set this period to fit your shift length, for example 8 hours. This
would enable you to prepare the instrument before the work shift starts and so avoid
having to interrupt sample processing for performing calibrations.
Precalibration
At any given time, there is only one accepted calibration result for each test. You can,
however, install and precalibrate one reserve reagent set. This is done, for example, to
ensure continuous sample processing.
Quality control (QC)
QC is performed at regular intervals to check the integrity of the whole measuring
system. For each test, up to three controls are defined. The results are compared
against predefined ranges or values and then interpreted accordingly.
Control
A control is a sample that has been measured using all tests it is associated with, in
order to define the ranges and values that determine the correct functioning of the
instrument. This is typically done both for the normal and the pathological analyte
concentration.
When QC is due
With regards to when it needs to be performed, QC is divided in the following types:
o
QC After Cal
The QC measurement is due after calibration of the test.
o
Interval QC
QC is due whenever its interval has expired. QC measurements of this type are
performed in a batch, typically once or twice a day.
o
Default QC
QC is performed at certain times during routine operation. This is done to fit in
with laboratory processes and procedures.
Ways of performing QC
With regards to how QC is performed, the following methods are provided:
o
Default QC
Default QC
is an automated process for performing multiple QC measurements
at the time when you define the QC orders. This is the ideal method if you want to
perform QC at certain times and days.
This method only applies to tests whose controls are defined to be performed as
part of
Default QC
. Therefore, if you intend to work with the
Default QC
function, you need to configure the tests accordingly.
Default QC
follows a streamlined procedure where the necessary QC orders are
automatically defined as soon as you identify a control. An order is defined for
each test for which this control is defined, provided the test is currently active on
the system. A wizard helps you select the controls, and a placement list supports
you in preparing and loading the controls.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...