7 Configuration
Cedex Bio System
Configuration
Short guide to extra wash cycle definitions
f
Utilities
>
Applications
>
Extra Wash Cycles
>
Extra wash cycle definitions
Item
Values
Comment
Active
On
,
Off
Defines whether the extra wash cycle can be applied
or not.
Action
None
Before
After
Defines when, in relation to the pipetting action from
the bottle defined in
Trigger Bottle Code
and
Pipet-
ting Type
, the extra wash cycle should be performed.
Trigger-Bottle
Code
0 ... 999
Bottle set code of the fluid that triggers performance
of the extra wash cycle.
If you use the pipetting type
Sample
, it defines the
bottle whose fluid will be prevented from being car-
ried over to the sample by using the extra wash cycle.
Pipetting Type
Reagent 1
Reagent 2
Start Reagent
Sample
Defines the individual bottle (bottle of the set or a
sample tube) whose pipetting triggers performance of
the extra wash cycle.
Cleaner-Bottle
Code
0 ... 999
Material code of the cleaner.
Volume
μ
L
Amount of cleaner that will be pipetted during the
extra wash cycle.
Table B-29
Extra wash cycle definitions
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...