6 Special operations
Replacing the probe
Replacing the probe
If the probe is bent, broken, or corroded, you need to replace it.
Tools and materials required
o
Probe and tubing set
o
Tube with ISE Deproteinizer
o
Tube with Activator
o
Glass beaker
a
To remove the probe
1
Make sure the system is in
Standby
status.
2
Shut down the system and switch off the instrument.
3
Remove the transfer head cover.
Press the release buttons on both sides and lift.
4
Release the tube leading to the probe from all tubing clips.
WARNING
Injury through working solutions
Direct contact with cleaning solutions or other working solutions may cause personal
injury. When handling such solutions, exercise the precautions required for handling them,
observe the cautions given in the package insert, and observe the information given in the
Safety Data Sheets available for Roche Diagnostics cleaning solutions.
Infection by biohazardous materials
Contact with samples containing material of human origin may result in infection. All mate-
rials and mechanical components associated with samples of human origin are potentially
biohazardous.
o
Be sure to wear protective equipment. Take extra care when working with protective
gloves; these can easily be pierced or cut, which can lead to infection.
o
If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
o
If waste solution contacts your skin, wash it off immediately with water and apply a dis-
infectant. Consult a physician.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...