6 Special operations
Importing data
Importing certificates
Certificates are digital records that ensure the authenticity of reagent barcodes. On
each
Cedex Bio
instrument there must be a certificate installed.
a
To import certificates
1
Choose
Utilities
>
Import
>
Certificate
.
A screen is displayed, that lists the currently installed certificates.
2
Press .
A screen is displayed, asking you to insert the USB stick.
3
Insert the USB stick.
4
Press .
5
Select the certificate.
You recognize a certificate by its file extension .prm.
6
Press
to confirm the selection.
The certificates are installed on the system.
The screen is displayed again, that lists the currently installed certificates.
Importing automatic mixing and extra wash cycle definitions
Mixing and extra wash cycle information is contained in a separate barcode on a
barcode transfer sheet.
a
To import mixing and extra wash cycle information
1
Choose
Utilities
>
Import
>
EWC/Mixing
.
A screen is displayed, asking you to scan the barcode.
2
Scan the barcode.
The data are installed on the system. A message will inform you when the
installation is complete. The screen for importing data is displayed again.
e
For information on defining extra wash cycles see
After the database installation
o
Load all bottle sets that are defined, using the appropriate software features.
o
Perform
Abs. Air/Water Calibration
maintenance action.
o
If
TRL Check
is activated, the value for the lower limit of the measuring range is
not
overwritten when importing the database. Check the relevant application documenta-
tion for the exact value, you may need to adjust it manually. See
on page B-142. (Note that there are applications where the lower limit of the measur-
ing range cannot be changed, e.g. ALBU2.)
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...