5 Daily operation
Finishing the shift
Checking the transmission of results
This step is only relevant if you work with a host system.
Before deleting the results, you must make sure they were properly transmitted to the
host.
a
To check for results that were not transmitted
1
Choose
Workplace
>
Result Review
.
A screen is displayed that contains all sample results of the day.
2
Choose
>
Not Sent to Host
.
3
Press .
A screen is displayed that contains the sample results that have not been
successfully transmitted to the host.
4
Select a result.
5
Press .
A screen for validating results is displayed.
6
Press
Retransmit
.
The list with the results that were not transmitted successfully is displayed again.
The result you just retransmitted should no longer be present.
7
Select and retransmit the remaining results as described in Steps 4 through 6.
8
Choose
>
Not sent to Host
.
The list should now be empty.
9
If the results were not successfully transmitted, contact the administrator of the
host computer.
Starting the End Shift wizard
The easiest way of performing the end of shift tasks is by following the
End Shift
wizard.
a
To start the End Shift wizard
1
Choose Overview >
End Shift
.
A screen is displayed for performing daily backup.
Results must be accepted before they can be transmitted to the host.
See also the information on
on page B-165, in particular on the
Auto
Accept
settings.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...