5 Daily operation
Cedex Bio System
Validating sample results
Rerunning tests
Rerun
Performing a rerun means reperforming the same test with a different predefined
dilution. Typically, you perform reruns if the result is outside the test range.
a
To rerun a test
1
Validate the result.
e
If the sample is still on board, see
To validate sample results of on-board samples
e
If the sample is no longer on board, see
To validate sample results from the results list
2
Press
Rerun
.
The system automatically creates a new order and selects the test. (The dilution
factor is part of the test definition and therefore automatically selected. It cannot
be changed.) It then performs the test.
In the results list, the result of the rerun is shown on a separate line:
WARNING
Incorrect results due to insufficient fluid
Insufficient fluid may lead to inaccurate pipetting and consequently to incorrect results.
Always fill the tubes with enough fluid that at least the defined dead volume of fluid is left
when pipetting is complete.
e
A
First measurement
B
Rerun result
-v-: Rerun with dilution.
Figure B-19
B
A
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...